• Home
  • GALLERY OF SERVICES
    • ANESTHESIA GALLERY
    • AUTOCLAVE GALLERY
    • DENTAL GALLERY
  • General INFORMATION
    • SERVICE INFORMATION - Q&A
    • GENERAL SERVICES
  • Anesthesia Information
    • ANESTHESIA CERTIFICATION
    • Anesthesia Guideliness / Repairs
  • Autoclave Information
    • Autoclave Certification
    • Autoclave GUIDELINES / REPAIRS
  • Dental Information
    • Dental certification
    • dental guidelines / repairs
  • Service Area
  • Board Compliance
  • CONTACT US
STASIS Veterinary Equipment Service
  • Home
  • GALLERY OF SERVICES
    • ANESTHESIA GALLERY
    • AUTOCLAVE GALLERY
    • DENTAL GALLERY
  • General INFORMATION
    • SERVICE INFORMATION - Q&A
    • GENERAL SERVICES
  • Anesthesia Information
    • ANESTHESIA CERTIFICATION
    • Anesthesia Guideliness / Repairs
  • Autoclave Information
    • Autoclave Certification
    • Autoclave GUIDELINES / REPAIRS
  • Dental Information
    • Dental certification
    • dental guidelines / repairs
  • Service Area
  • Board Compliance
  • CONTACT US
STASIS Veterinary Equipment Service

AUTOCLAVE GALLERY


AUTOCLAVE INVASIVE MAINTENANCE  (TUTTNAUER 2340M)
Machine has gone through years of no maintenance or service to the extend that functional operations have seized or significantly slowed  •  Machine would no longer drain  •  Switches no longer activating  •  Use of 'Ohm reader' on circuit board required to identify power usage intervals   •  Machine disassembly  •  Lubrication of hinges  •  Rethreading pipe seals   •  Invasive cleaning  •  Preventative parts were restored  •  Others parts were replaced  •  Chemical and vigorous cleaning to remove calcification  •  Grime and biofilm removed to restore stainless steel shine  •  Class 5 Integrator Chemical Testing  •  Biological Indicator (Spore Count) Testing  •  Multiple cycles needed to flush chemicals  •  Machine is ready for use again, hopefully for many more years to come
AUTOCLAVE INVASIVE MAINTENANCE (PELTON & CRANE OCR)
Machine has gone through years of no maintenance or service to the extend that functional operations have seized or significantly slowed  •  Machine would no longer drain  •  Switches no longer activating  •  Use of 'Ohm reader' on circuit board required to identify power usage intervals   •  Machine disassembly  •  Lubrication of hinges  •  Rethreading pipe seals   •  Invasive cleaning  •  Preventative parts were restored  •  Others parts were replaced  •  Chemical and vigorous cleaning to remove calcification  •  Grime and biofilm removed to restore stainless steel shine  •  Class 5 Integrator Chemical Testing  •  Biological Indicator (Spore Count) Testing  •  Multiple cycles needed to flush chemicals  •  Machine is ready for use again, hopefully for many more years to come
AUTOCLAVE INVASIVE MAINTENANCE (*OLD* TUTTNAUER 2340M)
Customer request for cleaning and update on mechanical functions  •  Preventative maintenance replaces broken or worn parts like filters, gaskets, and lids  •  Other components are restored to a 'like new' condition using all original materials  •  Complete restoration and calibration of system functions  •  Upon inspection, significant blockage was discovered due to a 'bug' and 'broken' materials blocking drain passage  •  Chemical and Biological Indicators all passed  •  Completed service report with all system timings presented to customer.
AUTOCLAVE INVASIVE MAINTENANCE (TUTTNAUER 2340M)
Customer request for cleaning and update on mechanical functions  •  Preventative maintenance replaces broken or worn parts like valves and gaskets  •  Other components are restored to a 'like new' condition using all original materials  •  Complete restoration and calibration of system functions  •  Upon inspection, significant blockage was discovered due to a 'hair' and 'debris' materials blocking drain passage  •  Chemical and Biological Indicators all passed  •  Completed service report with all system timings presented to customer.
AUTOCLAVE INVASIVE MAINTENANCE (*OLD* RITTER/MIDMARK M9)
Older unit customer had asked for a check-up on •  Machine has significant age and wear  •  Use of 'Ohm reader' on circuit board required to identify power usage intervals   •  Machine disassembly  •  Lubrication of hinges and latches  •  Rethreading pipe seals   •  Invasive cleaning  •  Preventative parts were restored  •  Others parts were replaced  •  Chemical and vigorous cleaning to remove calcification  •  Grime and biofilm removed to restore stainless steel shine  •  Class 5 Integrator Chemical Testing  •  Biological Indicator (Spore Count) Testing  •  Multiple cycles needed to flush chemicals  •  Machine is ready for use again, hopefully for many more years to come
AUTOCLAVE INVASIVE MAINTENANCE (*DISCONTINUED* OMEGA ST)
Old obsolete autoclave (sterilizer) that had not been used in years with parts that are no longer in production  •  Machine would intermittently 'short' and had years of build up from 'dirty' wraps and non-distilled water  •  Testing and exploration to diagnose issues  •  Use of 'Ohm reader' on circuit board required to identify shorts at high power usage intervals   •  Machine disassembly  •  Lubrication of hinges  •  Rethreading pipe seals   •  Invasive cleaning  •  Preventative parts were restored rather than replaced because items are no longer available  •  Chemical and vigorous cleaning to remove calcification and grime and restore stainless steel shine  •  Class 5 Integrator Testing  •  Multiple cycles needed to flush chemicals  •  Machine is ready for use again, hopefully for many more years to come